Density meter
Type: DWF
The DWF meter is suitable for density measurement of liquid products in pipelines. The current density is displayed in grams per liter or kg/m³ on the scale of the display unit. On request scale according to customer requirement
Details
Applications
Density meter for liquid media in pipelines.
Applicable in all branches of industry due to the use of
corrosion-resistant wetted parts.
Reliability
Robust, mechanical measuring method with magnetic measured value transmission clear 90° scale
Special features
- Inline density measurement of liquids
- Linear scale
- No bypass needed
- Wetted parts made of stainless steel
- Heating possible
- For the use in hazardous areas
Technical data
Body:
- Materials: Stainless steel, Hastelloy (others on request)
- Process connection: DN 25, ASME 1″,DN 50, ASME 2″
- Flanges according to EN 1092, ASME B16.5, DIN 2512 (others on request)
- Nominal pressure: PN16, ASME Cl 150 (others on request)
- Process temperature: -20°C to +150°C
- Ambient temperature: -20°C to +80°C
- Protection class: IP65/ 47 (EN60529)
- Flow rate: up to 10 m3/h
- Nominal sizes: DN25 or DN50
- Media temperatures: -20°C to +200°C
- Pressure ratings: up to PN16
- Density range medium: 700 g/l to 1900 g/l
- Measuring span: 50 g/l to 600 g/l
- Flow rate*: 2500 l/h / 5000 l/h 10000 l/h
*Reference conditions according to IEC 770, water at 20°C
Display:
Aluminum (stove enameled) or stainless steel
Electrical outputs: inductive limit switches,inductive limit switches safety engineering).
micro switch (others on request)
Ambient temperature: -40°C to +80°C (without limit switches)
-40°C to +80°C (with limit contacts)
- Power supply: 14 … 30 VDC
- Outputs: passive, galvanically isolated
- Analog: 4…20 mA
- Binary 1 and 2: Ui=30 V, Ii=20mA, Pi=100 mW
- Ambient temperature: -40°C up to +70°C
- Ingress protection: IP 20 (EN60529)
Certificates and approval
Explosion protection:
- Armature: BVS 03 ATEX H/B 112
- Indication unit: Explosion protection: DMT 00 ATEX E075
- Type of protection: II 2G EEX ia IIC T6
- CE-Marking: Explosion Protection Directive 2014/34/EU
Electromagnetic Compatibility. EMC-Directive
Communication
HART-protocol
HART-protocol and 2 NAMUR-switches
HART-protocol and 1 NAMUR-switch / pulse output
Profibus-PA
Foundation Fieldbus
Downloads